TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; APS 1 (ROLLER PUMP) Back to Search Results
Model Number 801041
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/29/2014
Event Type  malfunction  
Event Description
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, there was a roller pump display failure.The device was not changed out, as the customer continued to use the device.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per the clinical review on (b)(6) 2014: a "beltslip" message was observed on the local display of the sucker roller pump (refer to mdr 1828100-2014-01058).Later in the case a new issue surfaced with this pump.The local display darkened (went out), but the user was still able to adjust pump speed with the local speed knob or central control monitor (ccm) slider without issue.Pump information (speed or flow rate) was displayed on the ccm.The loss of display did not affect the ability to perform the suction process.The pump was removed from service after the procedure.The case was completed successfully, without delay and without associated blood loss.There was no harm observed.
 
Manufacturer Narrative
This complaint is related to mdr 1828100-2014-01058.Evaluation is in progress, but not yet concluded.The subsidiary's manufacturing engineering center (mec) could not duplicate the reported issue.This unit was repaired at the mec in august 2014.At this time, the service engineer replaced the display board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
APS 1 (ROLLER PUMP)
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4349507
MDR Text Key5204635
Report Number1828100-2014-01059
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801041
Device Catalogue Number801041
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/02/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/19/2014
Initial Date FDA Received12/11/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-