• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FRESENIUS MEDICAL CARE NORTH AMERICA CRIT-LINE BLOOD CHAMBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FRESENIUS MEDICAL CARE NORTH AMERICA CRIT-LINE BLOOD CHAMBER Back to Search Results
Catalog Number CL10021021
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/15/2014
Event Type  malfunction  
Event Description
A hemodialysis inpatient user facility has reported that during treatment a blood leak occurred.The leak was visually observed at the connection between the crit-line blood chamber and the dialyzer.Estimated blood loss was less than 250 ml's.The rn indicated she tightened the lines once but the lines must have loosened after the blood warmed the lines.Patient was able to complete treatment due to the leak occurring at the end of treatment.The patient had no adverse effects and no medical intervention was required.Sample has been discarded.
 
Manufacturer Narrative
Plant investigation has not yet been completed.A follow up report will be filed upon completion of the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRIT-LINE BLOOD CHAMBER
Manufacturer (Section D)
FRESENIUS MEDICAL CARE NORTH AMERICA
kaysville UT
Manufacturer (Section G)
HEMA METRICS DIVISION
695 n 900 w.
kaysville UT 84037
Manufacturer Contact
tanya taft, rn, cnor
920 winter st.
waltham, MA 02451-1457
7816999751
MDR Report Key4350173
MDR Text Key5144980
Report Number2937457-2014-03418
Device Sequence Number1
Product Code KOC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K935958
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCL10021021
Device Lot Number14031304
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/15/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
FRESENIUS 2008T HEMODIALYSIS MACHINE
Patient Age68 YR
Patient Weight78
-
-