|
Model Number 80-0728 |
Device Problems
Break (1069); Sticking (1597)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 12/01/2014 |
Event Type
Injury
|
Event Description
|
During a surgery to implant a distal radius plate, the surgeon was using a driver to install a screw.When the surgeon tried to remove the driver from the screw, it was stuck.The driver was pulled vertically from the screw and the driver tip broke off in the screw.The surgeon was unable to get the driver tip from the screw head so it was left implanted in the patient.
|
|
Manufacturer Narrative
|
A visual examination under magnification was performed.An oblique fracture pattern was found before the radius where the shaft transitions from the hex portion of the driver to the round shaft portion of the driver.An oblique fracture pattern likely indicates a side loading (bending), which was applied to hex tip.The distance between the flats of the hex was measured.The distances were found to be.0590,.0585, and.0585 which is within the tolerance of.0587 +.003/-.005.Conclusions - a conclusion cannot be drawn based on the lack of information concerning the circumstances of this driver break.
|
|
Search Alerts/Recalls
|
|
|