Result: the penumbra coil 400 was intact with the pusher assembly.The proximal end of the pusher assembly was kinked approximately 5.0 cm from the end.The pet lock was intact on the proximal end of the pusher assembly.The renegade hi-flow microcatheter (non-penumbra device) was ovalized approximately 10.5 cm from the distal tip.The outer diameter of the coil, distal detachment tip (ddt), and pusher assembly were measured and determined to be within specification.Conclusion: the complaint has been evaluated.The complaint indicates that the penumbra coil 400 was inserted through a renegade hi-flow catheter (non-penumbra device) and a strong resistance was felt.Evaluation of the returned product revealed that the distal shaft of the renegade hi-flow microcatheter was ovalized.If this ovalization was present when attempting to advance the coil, strong resistance, as described in the complaint, would be encountered.The kink in the proximal end of the pusher assembly was likely related to attempts to advance the coil against resistance.The penumbra coil 400 devices are 100% functionally tested and visually inspected for damage during process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|