• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP 5 (CP5); CARDIOPULMONARY BYPASS PUMP SPEED CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP 5 (CP5); CARDIOPULMONARY BYPASS PUMP SPEED CONTROL Back to Search Results
Model Number 60-02-60
Device Problem Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Patient involvement (2648)
Event Date 11/09/2014
Event Type  Other  
Event Description
Sorin group (b)(4) received a report that, after running for approx 9 hrs during a training session, the cp5 displayed an error message and re-booted itself.There was no pt involvement.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the sorin centrifugal pump.The incident occured in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow-up will be sent when the investigation is complete.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the sorin centrifugal pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that, after running for approximately 9 hours during a training session, the cp5 displayed an error message and re-booted itself.The investigation found no problems after a 24 hour test run with different interventions.The cp5 successfully completed all tests.The reported issue could not be reproduced.The hkr board was replaced as a precaution.Manufacturing records showed no nonconformities.Sorin group (b)(4) will continue to monitor the market for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CARDIOPULMONARY BYPASS PUMP SPEED CONTROL
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghst 25
munich D 809 39
GM  D 80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghst 25
munich D 80 939
GM   D 80939
Manufacturer Contact
cheri voorhees, mgr
14401 west 65th way
arvada, CO 80004
3034676527
MDR Report Key4356792
MDR Text Key15905345
Report Number1718850-2014-00476
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-60
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/19/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-