Model Number 60-02-60 |
Device Problem
Other (for use when an appropriate device code cannot be identified) (2203)
|
Patient Problem
No Patient involvement (2648)
|
Event Date 11/09/2014 |
Event Type
Other
|
Event Description
|
Sorin group (b)(4) received a report that, after running for approx 9 hrs during a training session, the cp5 displayed an error message and re-booted itself.There was no pt involvement.
|
|
Manufacturer Narrative
|
Sorin group (b)(4) manufactures the sorin centrifugal pump.The incident occured in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow-up will be sent when the investigation is complete.
|
|
Manufacturer Narrative
|
Sorin group (b)(4) manufactures the sorin centrifugal pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that, after running for approximately 9 hours during a training session, the cp5 displayed an error message and re-booted itself.The investigation found no problems after a 24 hour test run with different interventions.The cp5 successfully completed all tests.The reported issue could not be reproduced.The hkr board was replaced as a precaution.Manufacturing records showed no nonconformities.Sorin group (b)(4) will continue to monitor the market for trends.
|
|