The customer reported an uncontained clear fluid leak of approximately one (1) ml during each cycle from the bottom of the blood sampling valve (bsv) of the coulter hmx analyzer with autoloader.The leak was a mixture of blood, diluent, lyse, stain and clenz reagent, and was observed during startup.The customer stated the manual aspiration probe appeared to be bent, and the operator was unable to press the manual mode start lever as it would jam the probe.There was no report of injuries, erroneous results, direct physical contact with the leak, or change to patient treatment in connection with this event.The customer was wearing gloves and laboratory coat when the leak occurred.A beckman coulter (bec) field service engineer (fse) was dispatched to the customer site to evaluate the instrument.
|