Model Number M00565100 |
Device Problems
Material Deformation (2976); Positioning Problem (3009)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/05/2014 |
Event Type
Injury
|
Event Description
|
It was reported to boston scientific corporation that a wallflex enteral colonic stent was implanted in the rectum during a colon stent deployment procedure performed on (b)(6) 2014.According to the complainant, the stent was implanted to treat a malignant stricture as a bridge to surgery (bts).Reportedly, the patient's anatomy was tortuous.During the procedure, the physician deployed the stent, but noted on endoscopic image that a stent wire was protruding through the delivery system.The physician pulled the delivery system and the stent was released.The proximal wires tore and the physician then attached the deformed wires to the proximal end of the stent using clips.The physician noted that the stent was fully deployed and expanded but was not deployed at its intended location.The stent remains implanted.The physician reported there was no damage noted to the delivery system.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
|
|
Manufacturer Narrative
|
Patient's exact age is unknown; however, it was reported that the patient was over the age of 18.Reported event of stent placement / positioning problem.Reported event of stent broken.The complainant indicated that the device remains implanted and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
|
Search Alerts/Recalls
|
|
|