• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE OEC MEDICAL SYSTEMS (SLC) INSTATRAK 3500; RADIOLOGICAL IMAGE PROCESSING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE OEC MEDICAL SYSTEMS (SLC) INSTATRAK 3500; RADIOLOGICAL IMAGE PROCESSING Back to Search Results
Model Number INSTATRAK 3500
Device Problems Device Stops Intermittently (1599); Computer Operating System Problem (2898)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/03/2014
Event Type  malfunction  
Event Description
It was reported that the system would not boot up.The system was shut down, allowed to sit, then it would boot up.This caused an intermittent total loss of navigation functionality.There is no report of injury or death associated with the event described in this complaint.
 
Manufacturer Narrative
A ge service representative performed an on site investigation.The reported issue could not be duplicated.The system was tested and found to be working as intended and put back into service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSTATRAK 3500
Type of Device
RADIOLOGICAL IMAGE PROCESSING
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer (Section G)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer Contact
384 wright brothers dr.
salt lake city, UT 84116
8015364952
MDR Report Key4366057
MDR Text Key5287303
Report Number1720753-2014-09892
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberINSTATRAK 3500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/03/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-