PENUMBRA, INC. PENUMBRA COIL 400 DETACHMENT HANDLE; HCG, KRD Back to Search Results
Catalog Number DH1
Device Problem Separation Failure (2547)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/11/2014
Event Type  malfunction  
Event Description
The patient was undergoing a coil embolization procedure using a penumbra coil 400 detachment handle and a penumbra coil 400.During the procedure, the physician was unable to detach a coil although the detachment handle had broken the pet lock as intended.The coil was successfully removed and the procedure continued using additional coils.
 
Manufacturer Narrative
Conclusion: the product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This mdr is associated with mdr 3005168196-2014-00915.Device was disposed of by the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400 DETACHMENT HANDLE
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
1351 harbor bay parkway
alameda CA 94502
Manufacturer (Section G)
PENUMBRA, INC.
1351 harbor bay parkway
alameda CA 94502
Manufacturer Contact
kathleen kidd
1351 harbor bay parkway
alameda, CA 94502
5107483200
MDR Report Key4371972
MDR Text Key15913283
Report Number3005168196-2014-00916
Device Sequence Number1
Product Code HCG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120330
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 11/30/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date08/30/2017
Device Catalogue NumberDH1
Device Lot NumberF45556
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/20/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-