Result: the penumbra coil 400 was detached from the pusher assembly.The pet lock on the proximal end of the pusher assembly was intact.The proximal constraint sphere was still intact with the coil.The pull wire was inside the distal detachment tip (ddt).The ddt was not damaged.The inner diameter of the ddt was measured within specification.The outer diameter of the coil, proximal constraint sphere, pusher assembly, and the pull wire were measured within specification.Conclusion: the complaint has been evaluated.The complaint indicates that resistance was felt during the retraction of the coil, then it detach inside the px slim microcatheter.Evaluation of the returned product confirmed that the penumbra coil 400 was detached from the pusher assembly.This detachment might have occurred due to the force used during retraction of the coil and the resistance felt.The force use could have likely exceeded the compression strength of the ddt capture features mechanism.This is what keeps the coil intact with the pusher assembly.The outer diameter of the coil, pusher assembly, pull wire, and proximal constraint sphere were measured within specification.It is not clear if the px slim delivery microcatheter was damaged and caused resistance during the retraction of the coil.The px slim delivery microcatheter was not returned for evaluation.The root cause of this complaint cannot be determined.These devices are 100% functional tested during process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|