Model Number CD3231-40Q
Device Problem Failure to Interrogate (1332)
Patient Problems Dyspnea (1816); Low Blood Pressure/ Hypotension (1914); Cardiogenic Shock (2262)
Event Date 10/29/2014
Event Type  Injury  
Event Description
New information received notes that prior to replacement of the device the patient experienced dyspnea.
 
Manufacturer Narrative
Device evaluation anticipated, but not yet begun.
 
Manufacturer Narrative
The reported field event of an inability to interrogate the device was confirmed in the laboratory and was due to premature battery depletion.Based on all available parameter and usage information, device longevity was found to be below the expected limits.With an external power supply attached, the device functioned normally.No high current was detected during testing and the power consumption was normal.The original battery was returned to the vendor for further evaluation and analysis could not conclusively determine a root cause for the premature battery depletion.
 
Event Description
It was reported that the device could not be interrogated.All troubleshooting were unsuccessful in attempt to interrogate the device.The device was explanted and replaced.The patient exhibited symptoms of cardiogenic shock prior to the procedure, but was in stable condition following the device replacement.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D, DF-4 CONNECTOR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sharon mathew
645 almanor avenue
sunnyvale, CA 94085
4085226327
MDR Report Key4372235
MDR Text Key5440972
Report Number2938836-2014-19366
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2011
Device Model NumberCD3231-40Q
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/29/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
-
-