Model Number CD3367-40Q
Device Problems Failure to Interrogate (1332); Premature End-of-Life Indicator (1480)
Patient Problem Dizziness (2194)
Event Date 10/29/2014
Event Type  Injury  
Event Description
It was reported that the patient presented with dizziness.The device could not be interrogated and had reached eol.The device was explanted and replaced.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.Evaluation description: the reported field event of an inability to communicate with the device was confirmed in the laboratory and was due to low battery voltage.The low battery voltage was due to high current draw.The high current draw was caused by an anomalous component on the hybrid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA DR CRT-D, DF-4 CONNECTOR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sharon mathew
645 almanor avenue
sunnyvale, CA 94085
4085226327
MDR Report Key4372260
MDR Text Key5233718
Report Number2938836-2014-19338
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2015
Device Model NumberCD3367-40Q
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/29/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/30/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-