MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 97714
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem Therapeutic Effects, Unexpected (2099)
Event Type  malfunction  
Event Description
It was reported that the patient felt stimulation starting and stopping (stuttering) when using their adaptive stimulation programs.This occurred when the patient was standing or sitting perfectly still.The patient stated that it was unrelated to the 'stopping/starting' that he feels as the adaptive stimulation adjusts when he changed positions.It had not affected his therapy.The patient was concerned that it may indicate an issue with the stimulator and he worried it may suddenly stop on him.The company representative who saw the patient asked him to try the same programs without the adaptive stimulation on to see if this is replicated.Impedances were run and all came back within normal limits.Reprogramming was also done.The patient¿s status was alive with no injury.Additional information regarding further troubleshooting and outcome was requested.If received a follow up report will be sent.
 
Manufacturer Narrative
Concomitant products: product id 97740, serial # (b)(4), product type programmer, patient; product id 97754, serial # (b)(4), product type recharger; product id 977a260, serial # (b)(4), implanted: (b)(6) 2014, product type lead; product id 977a260, serial # (b)(4), implanted: (b)(6) 2014, product type lead.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4376007
MDR Text Key5284259
Report Number3004209178-2014-24752
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2015
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/10/2014
Date Device Manufactured06/05/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00058 YR
-
-