The hospital reported that, during a case, after placing the patient on their side, the patient was noted to move.The patient was reportedly administered propofol.The vaporizer was set at 7%, and the monitor reportedly read 3.2 %.The clinician reportedly increased the vaporizer dial setting and continued the case.There was no reported patient injury.
|
It should be noted that, according to the tec 7 user reference manual, the tec 7 is only to be mounted on a ge healthcare selectatec manifold.The vaporizer in the reported complaint was mounted on a nonvalidated spacelabs anesthesia machine.A third party service representative performed an output check of the vaporizer and found low output.The vaporizer was returned to the ge healthcare manufacturing site for investigation.The unit was tested, and the output was found to be within published specifications.
|