• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, INC. SYNCARDIA COMPANION EXTERNAL LI-BATTERY PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, INC. SYNCARDIA COMPANION EXTERNAL LI-BATTERY PACK Back to Search Results
Catalog Number 293001-001
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 12/21/2014
Event Type  malfunction  
Event Description
This companion external battery was not in use by a patient.The reported issue involves syncardia companion external batteries and are reported under two separate medical device reports: companion external battery s/n (b)(4); and companion external battery s/n (b)(4) (mfr report # 3003761017-2014-00325).This alleged failure mode poses a low risk to a patient because the issue was observed when the companion external batteries were not in patient use.In addition, this alleged failure mode would not prevent a companion 2 driver from performing its life-sustaining functions.The companion 2 driver has multiple power sources of external wall power, external batteries and an internal emergency battery.
 
Manufacturer Narrative
The results of the investigation will be provided in a supplemental mdr.
 
Manufacturer Narrative
(b)(4).
 
Event Description
This companion external battery was not in use by a patient.The reported issue involves syncardia companion external batteries and are reported under two separate medical device reports: (1) companion external battery s/n (b)(4) (mfr report # 3003761017-2014-00324); and (2) companion external battery s/n (b)(4) (mfr report # 3003761017-2014-00325).The customer reported that the companion 2 external batteries are "dead" and need to be returned.The companion 2 external batteries were returned to syncardia for evaluation.Review of the external components of companion 2 external battery s/n (b)(4) and s/n (b)(4) did not reveal any abnormalities.Testing of companion 2 external battery s/n (b)(4) revealed that the battery did not display output voltage, did not accept charge current and the gas gauge did not indicate charge when the button was pressed.The failure of the battery to operate during testing confirmed the customer reported issue.The root cause of the inoperable battery is unknown, but the results observed are consistent with the lack of use and charging.Testing of companion 2 external battery s/n (b)(4) revealed that the battery was able to establish communication with testing software when connected to an external power source.During testing, it was observed that the battery was in a state of significant over-discharge.However, once the battery was connected to an external power source, the battery was able to enter into "fast charge" mode and fully recharge.In addition, the battery sat for an additional 48 hour rest period.Following the rest period, the battery was verified to have retained the full state of charge and no self-discharge condition existed.The root cause for the customer reported experience is that the cells of companion 2 external battery s/n (b)(4) were in a state of over-discharge and would not enter into "fast charge" mode as required by companion 2 driver firmware.Companion battery s/n (b)(4) was taken out of service.Companion battery s/n (b)(4) was serviced and passed all final performance testing.This alleged failure mode poses a low risk to a patient because the issue was observed when the companion external batteries were not in patient use.In addition, this alleged failure mode would not prevent a companion 2 driver from performing its life-sustaining functions.The companion 2 driver has multiple power sources of external wall power, external batteries and an internal emergency battery.This issue will continue to monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION EXTERNAL LI-BATTERY PACK
Type of Device
BATTERY PACK
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key4380567
MDR Text Key5117017
Report Number3003761017-2014-00324
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor,distributor,foreign,hea
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number293001-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-