• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Pumping Stopped (1503)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/01/2014
Event Type  Other  
Event Description
Sorin group received a report that the s5 roller pump stopped and alarmed during a procedure.The pump was changed out and the case was completed without further issues.There was no report of patient injury.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the s5 roller pump stopped and alarmed during a procedure.The pump was changed out and the case was completed without further issues.A sorin service representative was dispatched to investigate.The reported problem could not be reproduced but the pump had warning triangles in rpm and lpm sections of the display and the pump would not turn.The biomedical engineer had replaced the touchscreen, making the pump operational.It was found that the cable for the can was in pump b instead of the can on the processor.This was corrected and all testing showed no problems.The system was also updated.No further problems have been reported since these actions.No nonconformities were noted during manufacturing record review.The issue will be monitored for trends and if identified, corrections will be recommended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
cheri voorhees, manager
14401 west 65th way
arvada, CO 80004
3034676527
MDR Report Key4387932
MDR Text Key5436800
Report Number1718850-2014-00486
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/24/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-