• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES; DEFIBRILLATORS, AUTOMATIC, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES; DEFIBRILLATORS, AUTOMATIC, EXTERNAL Back to Search Results
Catalog Number VLP12-02B
Device Problem Material Fragmentation (1261)
Patient Problem No Patient Involvement (2645)
Event Date 12/15/2014
Event Type  malfunction  
Event Description
The customer contacted physio-control to report that pins had broken off of a hard paddle assembly and become lodged in the device's therapy connector.As a result, defibrillation may not be possible.There was no patient use associated with the reported event.
 
Manufacturer Narrative
(b)(4): physio-control provided the customer, a biomedical engineer, with technical assistance.It was later confirmed by the biomedical engineer that he replaced the device's therapy connector as well as the hard paddles assembly which resolved the reported issue.After observing proper device operation through functional and performance testing the unit was placed back into service for use.Neither the device, nor the hard paddles, have been returned to physio-control for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES
Type of Device
DEFIBRILLATORS, AUTOMATIC, EXTERNAL
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4388524
MDR Text Key16541957
Report Number3015876-2015-00017
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102972
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberVLP12-02B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age12 YR
Event Location Hospital
Date Manufacturer Received12/15/2014
Date Device Manufactured10/24/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-