The patient was undergoing a thrombectomy procedure using a penumbra system 5max ace reperfusion catheter.During the procedure, a hole was noticed in the catheter.The physician attempted to occlude the hole; however, was unsuccessful.A new 5max ace was used to complete the procedure.There was no report of an adverse effect on the patient.
|
Result: the penumbra reperfusion catheter 5max ace was returned with tape and plastic wrapped around the strain relief covering the damage.The catheter was kinked approximately 2.5 cm from the hub.Conclusion: the complaint has been evaluated.The complaint indicated that the penumbra reperfusion catheter 5max ace had a hole near the hub.Evaluation of the returned device confirmed a kink in the shaft of the catheter underneath the strain relief.This type of damage typically occurs when the product is improperly handled during removal from the packaging hoop.If the product is removed at angle while force is being applied, it is likely that damage will occur.These devices are 100% visually inspected for damage during process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|