PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DEFIBRILLATORS, AUTOMATIC, EXTERNAL Back to Search Results
Model Number 20
Device Problem Material Fragmentation (1261)
Patient Problem No Patient Involvement (2645)
Event Date 12/16/2014
Event Type  malfunction  
Event Description
The customer contacted physio-control to report that a pin from the defibrillation hard paddle assembly had broken off inside the therapy connector assembly of the device.As a result, defibrillation therapy would not likely be available because another defibrillation therapy cable or hard paddle assembly would not plug in to the device.There was no patient use associated with the reported event.
 
Manufacturer Narrative
(b)(4): the customer ordered a replacement therapy connector assembly and hard paddle assembly to repair their device.After observing proper device through functional and performance testing, the device will be returned to use.The device and removed parts were not returned to physio-control for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DEFIBRILLATORS, AUTOMATIC, EXTERNAL
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4403435
MDR Text Key5123934
Report Number3015876-2015-00027
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age9 YR
Event Location Hospital
Date Manufacturer Received12/16/2014
Date Device Manufactured05/25/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-