• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number ER420
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/18/2014
Event Type  malfunction  
Event Description
It was reported that prior to an unknown procedure, both devices the clips bifurcated during the first trial fire, after taken out of the package.It is unknown what was used to complete the procedure.There were no adverse consequences for the patient reported.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
(b)(4).Additional information: the analysis results found that the er420 device was returned in good condition with a clip in the jaws; the clip was removed in order to inspect the jaws and they were found to be in good condition.In an attempt to replicate the reported incident, the device was tested for functionality.During the analysis, the device was cycled and it fed, retained and formed the remaining clips as intended.The reported event could not be confirmed as the device was found to be fully functional.No conclusion could be reached as to what may have caused the reported incident.The batch record was reviewed and no anomalies were noted during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGACLIP ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 0096 9
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 0096 9
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4405260
MDR Text Key5433811
Report Number3005075853-2015-00274
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/20/2019
Device Catalogue NumberER420
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/21/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-