• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR XP; IMMUNOASSAY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR XP; IMMUNOASSAY ANALYZER Back to Search Results
Model Number ADVIA CENTAUR XP
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/16/2014
Event Type  malfunction  
Event Description
A discordant, falsely elevated troponin result was obtained on a redraw sample (sample id: (b)(6)) of one patient on an advia centaur xp instrument.The discordant result was not reported to the physician(s), as a previous sample obtained from the patient (b)(6) was questioned and resulted lower on multiple repeats.The redraw sample was repeated four times on an advia centaur instrument, resulting lower.The corrected result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated troponin result.
 
Manufacturer Narrative
The customer contacted the siemens customer care center (ccc).Quality controls were within range at the time of event.The cause of the discordant, falsely elevated troponin result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR XP
Type of Device
IMMUNOASSAY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS
manufacturing limited
registration number: 8020888
chapel lane, swords, co., dublin
EI  
Manufacturer Contact
cassandra kocsis
511 benedict avenue
tarrytown 10591
9145242687
MDR Report Key4405472
MDR Text Key5431628
Report Number2432235-2015-00018
Device Sequence Number1
Product Code MOI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA CENTAUR XP
Device Catalogue Number078-A011-03
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/16/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/28/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-