Complaint received reporting leakage incident with use of 20123-01 5" bag spike adaptor w/ spiros, vented cap and unk hosp pump tubing set.The info received reports vp-16 tubing leaked at connector (site) ".As rn connected the flush bag to the tubing.Remaining vp-16 (<5 cc) caught by rn with gloves.Chemo fell down gown sleeve onto rn's skin and long sleeve shirt.Blood began to return through tubing still connected - rn clamping tubing, disconnected it, and flushed line.Rn washed site (l forearm) thoroughly." following chemo exposure treatment protocol, clinician reportedly experienced no adverse consequences.Device return: one (1) used 20123-01 5" bag spike adapter w/ spiros, attached to an unk ext.Set.Visual inspection (pre and post decontamination) was performed.The report noted that during decontamination flushing, leakage was seen originating from the male end of the 20123-01 device sets spiros connector.Upon closer inspection damages were seen on the body of the spiros connector.
|