• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; HEATED HUMIDIFICATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; HEATED HUMIDIFICATION SYSTEM Back to Search Results
Catalog Number 425-00
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/16/2014
Event Type  malfunction  
Event Description
The customer alleges that there is a fault with the blue pigtail of the device.The unit faulted as soon as the blue pigtail was connected into the circuit no patient injury.The condition of the patient is reported as fine.
 
Manufacturer Narrative
(b)(4).A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned.The dhr (device history record) investigation did not show issues related to complaint.Serial #(b)(4) was manufactured on 06/29/2007.A document assessment was conducted and no changes were required.Corrective actions cannot be established since it is necessary to receive the physical sample to perform a proper investigation and confirm the defect.At this time due to that the device sample is not available it is not possible to determine the source of the defect reported.Corrective actions cannot be established since it is necessary to receive the physical sample to perform a proper investigation and confirm the defect.If the device sample becomes available at a later date, this complaint will be updated accordingly.Teleflex will continue to monitor and trend similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HEATED HUMIDIFICATION SYSTEM
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate, b.c. 2147 8
MX   21478
Manufacturer Contact
margie burton, rn
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key4412067
MDR Text Key5372623
Report Number3003898360-2015-00007
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number425-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received12/16/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
CIRCUIT
-
-