The customer contacted the siemens customer care center (ccc).Ccc evaluated the instrument data and discovered incorrect purge settings in centralink.The purge command was set to purge samples with "pending" status.The samples affected in this event were in "scheduled" status, and therefore, were not purged by centralink.Ccc implemented an additional "omit pending samples" command, by remote connection, to include "scheduled", "rerun" and "review" statuses.The cause of incorrect patient identification numbers obtained for two patient samples was an incorrect configuration.The instrument is performing according to specifications.No further evaluation of the device is required.
|