(b)(4).All available information has been provided to the actual manufacturer, b.Braun (b)(4).Their report states that they received two samples.For sample #1, the safety clip was activated and engaged properly on the tip of cannula.No clip deformation was observed.The used sample was checked by repositioning the clip and retesting its function and found that the clip engaged properly onto the tip of the needle.There was no rough surface/dented mark observed on cannula surface.For sample #2, the safety clip was not in engaged position and was dislodged at the middle of cannula.The used sample was checked by repositioning the clip and retesting its function and found that the clip engaged properly onto the tip of the needle.There was no abnormality found while pulling the catheter out from the cannula.The clip engaged properly onto the tip of the needle.There was no rough surface/dented mark observed on cannula surface.The returned sample #1 is g18 and returned sample #2 is g22.Based on the above test, there was no defects on the safety clips.Moreover, the assembly machine is also equipped with a vision system which conducts 100% inspection on the position of the clip in the hub.The device history record was reviewed and there were no such defect encountered during in-process inspection.Process cards show no abnormalities.If additional pertinent information becomes available, a follow report will be submitted.(b)(4).
|