Upon receipt, the device could not be interrogated, confirming the clinical observation.Therefore, the device was opened to inspect the inner assembly and to check the functionality of the electronic module.During the analysis of the electronic module, it was revealed that the output stage of one high voltage circuit and several other electric components had been damaged.These damages led to an elevated current consumption, which depleted the battery.Due to this and the damages of the electronic module the device could not be interrogated properly.Based on the damage symptoms of the electronic module, the damages were most probably caused by the reported external defibrillation.In general the icd is protected against the high voltage discharge during external defibrillation, but depending on the position of the external defibrillation electrodes and individual patient circumstances a damage of the electronic module cannot be excluded.In addition the manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process which might be related to the observed damage symptoms.Particularly the final acceptance test proved the device functions to be flawless.There was no indication of a material or manufacturing problem.
|