The customer reported that the quality controls for partial pressure of oxygen (po2) were running consistently below the mean.When running patient samples, it was determined that the analyzer was generating lower po2 patient results when compared to another cobas b 123 analyzer.Data was provided for a total of three patient samples.Upon evaluation it was determined that the questioned po2 results were flagged, alerting the user there was a problem with the results.However, it was determined that two of the patient samples had erroneous base excess of blood (be) and bicarbonate concentration in plasma (chco3) results.It was asked, but it is not known if any erroneous results were reported to parties outside of the laboratory.The first patient sample had an initial be result of 6.24 mmol/l.The same sample was repeated on the same analyzer, resulting as 5.50 mmol/l.It was stated that the results obtained on this analyzer were not consistent with the symptoms and diagnostics of the patient in question.The same sample was tested on a different cobas b 123 analyzer and resulted as 4.6 mmol/l.The results from the different cobas b 123 analyzer were stated to be consistent with the patient's symptoms.The second patient sample, from a female born on (b)(6) 1949, had initial results of 9.31 mmol/l for be and 34.9 mmol/l for chco3 on (b)(6) 2015.The sample was repeated on a cobas b 221 analyzer, resulting as 6.2 mmol/l for be and 31.6 mmol/l for chco3 on (b)(6) 2015.It was asked, but it is not known if the initial and repeat results were from the same sample or if each was measured from a different sample collection.The patients were not adversely affected.From (b)(6) 2014 to (b)(6) 2015, sensor cartridge lot 21543081 with an expiration date of january 2015 was used.From (b)(6) 2015, sensor cartridge lot 21543782 with an expiration date of march 2015 was used.
|