• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GETINGE DISINFECTION AB GETINGE; WASHER DISINFECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GETINGE DISINFECTION AB GETINGE; WASHER DISINFECTOR Back to Search Results
Model Number 46-5
Device Problem Residue After Decontamination (2325)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/08/2014
Event Type  malfunction  
Event Description
A small piece of blue debris/substance was observed in a phacoemulsification hand piece after cleaning in a 46 series washer disinfector using a rack accessory (mis uo-module_mis modular - l shape with filter).This was observed 5-6 times at this facility.
 
Manufacturer Narrative
The specific model number of the washer disinfector 46-5t is not available in the us.However, the accessory rack that may be the potential source of the blue debris is utilized for all 46 series washer disinfectors that are available in the us.F/u report to be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GETINGE
Type of Device
WASHER DISINFECTOR
Manufacturer (Section D)
GETINGE DISINFECTION AB
ljungadalsgatan 11 (se-07)
vaxjo 351 15
SW  351 15
Manufacturer Contact
ljungadalsgatan 11
(se-07)
vaxjo 351 1-5
2725036
MDR Report Key4458960
MDR Text Key17579163
Report Number9616031-2014-00005
Device Sequence Number1
Product Code MEC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number46-5
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-