MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD NEOPUFF INFANT RESUSCITATOR; BTL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD NEOPUFF INFANT RESUSCITATOR; BTL Back to Search Results
Model Number RD900
Device Problems Break (1069); Medical Gas Supply Problem (2985)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
A hospital in (b)(6) reported that an rd900 neopuff infant resuscitator had a broken gas outlet connector and requested a repair.No patient consequence was reported.
 
Manufacturer Narrative
(b)(4).Method: the complaint neopuff was received at our regional office in (b)(6), where it was inspected by a trained fisher & paykel healthcare service engineer.Our investigation is based on photographs and information provided by our (b)(4) office.Results: visual inspection of the photographs of the subject neopuff revealed that the gas outlet port and the upper end cap were cracked.A lot check revealed no other complaints of this nature for lot 090317.Conclusion: the nature of the cracking suggests that the device has sustained some form of impact damage.The neopuff is a portable, reusable device used to assist in the delivery of respiratory breaths to infants until adequate spontaneous breathing occurs.Being a portable device, the neopuff can be susceptible to impact damage, for instance when accidentally dropped or subjected to considerable external force.The neopuff technical manual warns against dropping the neopuff or subjecting it to impact damage which may cause the unit to operate incorrectly.If the neopuff is suspected to have been damaged, the manometer and valve system should be performance tested.The neopuff technical manual states the following: "dropping the neopuff / perivent infant resuscitator or other similar forms of impact may cause damage resulting in incorrect operation of the unit.If you suspect damage to have occurred, please perform checks as outlined [in the manual] before connection to a patient".The neopuff was repaired by the (b)(6) fisher & paykel healthcare service centre and returned to the customer after passing the safety and performance checks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOPUFF INFANT RESUSCITATOR
Type of Device
BTL
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
15365 barranca parkway
irvine, CA 92618-2216
9194534000
MDR Report Key4472543
MDR Text Key19989565
Report Number9611451-2015-00039
Device Sequence Number1
Product Code BTL
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K892885
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRD900
Device Catalogue NumberRD900
Device Lot Number090317
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/20/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/17/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-