It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the flow sensor on centrifugal pump was not reading values.The device was not changed out.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per the clinical review on (b)(6) 2015: the perfusionist (ccp) stated there were no issues during set-up and priming and for most of the cpb period.During the last minutes of cpb, when the patient was being weaned from cardiopulmonary bypass, the flow display on the sarns centrifugal unit was displayed as 0.00.The ccp stated the motor continued to work and there were clinical indications that there was blood flow in the circuit and patient.The ccp disconnected the sensor from the tubing and placed some ultrasonic gel on the sensor and re-connected the sensor to the tubing.The flow continued to be displayed as 0.00.The ccp disconnected the flow sensor from the control module and re-connected the sensor and 0.00 continued to be displayed.The ccp weaned the patient off cpb without difficulty.While off cpb, the ccp found an older level sensor in storage and connected the sensor to the control module and with the motor off and lines clamped the sensor was reading very erratic flows (according to the ccp bouncing all over) even though there was no actual flow.The control module and sensors were removed from service and the manufacturer was notified.The case was completed successfully with no delay and no associated blood loss.There was no harm observed.
|