• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS INC PULSE GEN MODEL 102; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS INC PULSE GEN MODEL 102; GENERATOR Back to Search Results
Model Number 102
Device Problems Premature Discharge of Battery (1057); Leak/Splash (1354); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/13/2005
Event Type  malfunction  
Event Description
It was reported that the patient had generator replacement on (b)(6) 2014 due to neos=yes according to the referring nurse practitioner.Analysis of the generator was completed on (b)(6) 2015.The elective replacement indicator flag was set and an open can measurement of the battery voltage confirmed that the eri flag had been properly set; the battery was operating at low voltage (depleted level).With the exception of the eri parameter (eri flag was set to ¿yes¿ due to a low battery condition), the device performed according to functional specifications.It was identified that there was increased current consumption for the device which potentially contributed to a premature end of the battery life.Analysis on the test bench identified and confirmed that c6 had excessive leakage current.With the capacitor substitution for c6, the pulse generator module performed according to the functional specifications.The most probable root cause for the premature end of service and low battery condition was identified to be a leaky capacitor, c6.
 
Manufacturer Narrative
Device failure occurred, but did not cause or contribute to a death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 102
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4480816
MDR Text Key20788729
Report Number1644487-2015-03740
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/30/2007
Device Model Number102
Device Lot Number012830
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2014
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received01/29/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age44 YR
-
-