The hill-rom technician found all four casters rubber wheels are worn out from wear and tear causing bed to slightly move.Per the hill-rom service manual, the affinity three birthing bed and affinity four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventative maintenance.Check the tires for cuts, wear, tread life, etc.Apply the brake, and check to ensure that the bed will not move.If the bed moves, inspect it for wear, and adjust if required.Apply the steering pedal and check the steering to ensure proper locking action when activated.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2012-2014.It is unk if the facility performed any other preventative maintenance on this bed.The technician replaced all four brake casters to resolve the issue.Based on this info, no further action is required.
|