The echelon is a 1.5 tesla closed bore mri system.From (b)(6) 2014, hitachi dispatched a field service engineer to investigate the incident.The spine receive coil and transmitter were heat tested and the tests were negative.The ctl temp was 75 deg f, the transmitter 80 deg f.The transmitter tuning was checked and found to be out of specification.A bad tuning capacitor was found and replaced.The transmitter was retuned and is now properly aligned.Even though the transmitter was mistuned, the transmitter gain was in the normal range, so rf power output was within specification.Service also found that one of the gantry bore fans was partially blocked by lint.The last preventive maintenance visit where the fans were checked was in (b)(6) 2014.(pms are quarterly.) clinical applications reviewed the patient images and found good image quality and no evidence of a malfunction.Sar values ranged from 0.97 to 2.21 w/lg.Total scan time was approximately 27-28 minutes each for a t and l spine exam.Room temp and humidity were within specification.Evidence suggests that, due to the patient's size, the body was close to the transmitter coil surface (bore) and the partially blocked fan reduced airflow.There was no evidence of skin to skin contact (causing rf current loop) but we cannot rule it out.
|