During a regularly scheduled preventative maintenance, physio-control observed that the customer's monophasic aed device would constantly alarm, stating that motion was detected, when attached to a patient simulator.As a result, defibrillation would not be possible because a patient rhythm could not be analyzed.There was no patient use associated with the reported event.
|
(b)(4).Physio-control examined the customer's device and verified the reported issue.Physio advised the customer that their monophasic aed device is no longer supported by physio and, as a result, the device is no longer serviceable.Physio recommended that the device be permanently removed from service and a replacement be obtained.The cause of the reported issue could not be determined.
|