MEDTRONIC, INC. ATTAIN CLARITY; CATHETER, FLOW DIRECTED Back to Search Results
Model Number 6225I
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/27/2014
Event Type  malfunction  
Event Description
It was reported that the during the procedure the balloon catheter was obstructive in producing clear veinogram, occludes posterior vessels and unable to be positioned in suitable location to get diagnostic information needed.The balloon catheter was removed and replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN CLARITY
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4497973
MDR Text Key5446141
Report Number2182208-2015-00581
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/27/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6225I
Device Catalogue Number6225I
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/27/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-