• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI ATTAIN ABILITY; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ATTAIN ABILITY; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 419688
Device Problems Collapse (1099); High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2014
Event Type  Injury  
Event Description
It was reported that the left ventricular (lv) lead had high impedance.It was noted that the patient is prone to subclavian crushes.The lead was capped and replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Concomitant products: 3830-59 lead implanted: (b)(6) 2011.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN ABILITY
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4499999
MDR Text Key5532437
Report Number2649622-2015-01144
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/24/2011
Device Model Number419688
Device Catalogue Number419688
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/14/2014
Date Device Manufactured04/05/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
6947M-62 LEAD, DTBA1DA ICD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age00046 YR
-
-