|
Catalog Number 530.200 |
Device Problems
Overheating of Device (1437); Power Problem (3010)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Event Description
|
It was reported from (b)(6) that the battery device did not provide power to the device and heated up.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
(b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device.A functional assessment was performed and the reported condition was not confirmed or duplicated.Therefore, an assignable root cause was not determined.If information is obtained that was not available, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|