On (b)(6) 2015, the patient underwent endovascular repair of an abdominal aortic aneurysm using gore® excluder® aaa endoprostheses.After the implantation of trunk ipsilateral leg component, the delivery catheter of the plc201200j was inserted from the contralateral side through gore® dryseal sheath (dsl1428j/13048885).The device was successfully implanted, and when the delivery catheter was withdrawn from the patient, bleeding was observed from the valve of the sheath.The physician quickly inserted the dilator and the replaced the sheath with a spare.The transfusion was performed due to the blood loss (the amount unknown).The rest of the procedure was performed with no other issues.The patient tolerated the procedure.The physician reportedly suspected the damage to the valve occurred during the procedure.But the cause of the damage was unknown.
|