• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Disconnection (1171)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/15/2015
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the 3-way stopcock on the sheath disconnected from the sheath.This occurred when the physician was attempting to remove irrigation line connection with needle forceps from side port of the sheath.The stopcock popped off and fell on the floor.The physician quickly press fit another stopcock in its place.The procedure continued with same sheath without further incident.There were no patient symptoms or complications related to the event.
 
Manufacturer Narrative
The device has not yet been analyzed.Results of evaluation of returned device will be submitted in a supplemental report.
 
Manufacturer Narrative
The returned device was visually inspected and functionally tested.The reported issue has been confirmed through testing.The device failed the returned product inspection due to detached stopcock from sideport tube and post-procedure shaft kink.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland,qc H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland,qc H9H 5 H3
CA   H9H 5H3
Manufacturer Contact
voula radiotis
16771 chemin ste-marie
kirkland,qc H9H 5-H3
CA   H9H 5H3
5146941212
MDR Report Key4512811
MDR Text Key5533842
Report Number3002648230-2015-00034
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/01/2016
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number08087
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/12/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-