• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE OEC MEDICAL SYSTEMS (SLC) 9900; FLUOROSCOPIC X-RAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE OEC MEDICAL SYSTEMS (SLC) 9900; FLUOROSCOPIC X-RAY Back to Search Results
Model Number 9900
Device Problems Failure to Charge (1085); Poor Quality Image (1408)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/19/2014
Event Type  malfunction  
Manufacturer Narrative
A ge service representative performed an on site investigation.The battery charger board, filament driver pcb, power cap module and batteries were evaluated and replaced.The system was tested and found to be working as intended and returned to service.
 
Event Description
It was reported that the system had charger failed errors and would not create x-rays.This can cause the system to become unusable due to the inability to create a live fluoroscopic image.There is no report of injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9900
Type of Device
FLUOROSCOPIC X-RAY
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer (Section G)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer Contact
384 wright brothers dr.
salt lake city, UT 84116
8015364952
MDR Report Key4515731
MDR Text Key21847868
Report Number1720753-2014-08470
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9900
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/19/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-