• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE REHABILITATION EQUIPMENT CO. PERFECTO2 V WITH SENSOR 9153650799; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE REHABILITATION EQUIPMENT CO. PERFECTO2 V WITH SENSOR 9153650799; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number IRC5PO2V
Device Problems Loose or Intermittent Connection (1371); Failure to Power Up (1476); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Independent repair center: customer alleged alarming/red light and the unit will not start and the key failure is the hose clamp caused the heat exchanger to come loose from the manifold, so the hose clamp was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFECTO2 V WITH SENSOR 9153650799
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
INVACARE REHABILITATION EQUIPMENT CO.
no.5 weixi road, sip
suzhou jiangsu 2151 21
CH  215121
Manufacturer (Section G)
INVACARE REHABILITATION EQUIPMENT CO.
no.5 weixi road, sip
suzhou jiangsu 2151 21
CH   215121
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44036
8003336900
MDR Report Key4516531
MDR Text Key22061540
Report Number3008262382-2015-00581
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 02/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberIRC5PO2V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/06/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-