• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO UNIVERSAL HANDSWITCH; UNIT, OPERATIVE DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO UNIVERSAL HANDSWITCH; UNIT, OPERATIVE DENTAL Back to Search Results
Catalog Number 5100009000
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 01/20/2015
Event Type  malfunction  
Event Description
The universal handswitch was returned for service.Upon evaluation at the manufacturer, it was discovered that it would not lock in the run or safe position creating the potential for the device to be activated unintentionally.There was no patient involvement, and there were no user injuries or adverse consequences.
 
Manufacturer Narrative
(b)(4).The device failed for wear with use over time.The handswitch is not a repairable device and will not be returned to the user facility.
 
Event Description
The universal handswitch was returned for service.Upon evaluation at the manufacturer, it was discovered that it would not lock in the run or safe position creating the potential for the device to be activated unintentionally.There was no patient involvement, and there were no user injuries or adverse consequences.
 
Manufacturer Narrative
A follow up report will be filed after the quality investigation has been completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL HANDSWITCH
Type of Device
UNIT, OPERATIVE DENTAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key4520918
MDR Text Key5581893
Report Number0001811755-2015-00541
Device Sequence Number1
Product Code EIA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032117
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5100009000
Device Lot Number05179
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received04/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-