After the event a maquet field service technician (fst) was on site to investigate, the affected operation room table top.He found one defective electric drive.As a result the motor driven functions of the basic table top failed to move.However, in such a case all movements of the column and of the joint modules remain available.Also the table top transfer from the column onto the transporter can still be carried out.The operation room table system magnus is equipped with an emergency operation mode to obtain the functionality of the operation room table system as much as possible.The emergency operating mode is described in the instructions for use (b)(4).It is an isolated case.We are not aware of any similar incident.Maquet (b)(4) provides product failure investigation, analysis and resolution for the device described in this report.
|