• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310R
Device Problems Degraded (1153); Volume Accuracy Problem (1675)
Patient Problem No Patient Involvement (2645)
Event Date 01/22/2015
Event Type  malfunction  
Event Description
During evaluation of a returned homechoice device, a baxter technician determined the device failed fluid volume accuracy testing.No additional information is available.
 
Manufacturer Narrative
(b)(4) the homechoice device was returned and evaluated by the product analysis lab (pal).A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.An external/internal inspection was performed with no issues noted.Temperature verification testing was performed and the results were found to be within specifications.The device was determined to meet electrical performance specification requirements per rite (returned instrument test evaluation) testing.The device failed rite functional testing as fluid was transferred above the allowable specification range during volumetric accuracy testing.Upon further inspection, it was determined that the cause of the failure was due to deteriorated piston foam.The piston foam was to be scrapped and the device was sent for servicing.A capa has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4528246
MDR Text Key5544187
Report Number1416980-2015-06314
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C8310R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2015
Date Manufacturer Received01/22/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-