During a gastroscopic polypectomy, the customer attempted to push the slider of the subject device to release the loop after ligating the polyp but the slider couldn't work.The customer forcibly pulled the subject device with the polyp ligated, and removed the polyp physically.After removing the polyp, the procedure was completed using a hemostasis clip as preventive measures.No pt injury was confirmed by the scopic investigation the next day.
|
The subject device was returned to olympus for investigation.It is confirmed that the loop wasn't able to be released since the operation pipe of the subject device broke and the slider wasn't able to work.Ant it is supposed that the operation pipe was kinked, given tension, and broke.The mfg record for the same lot number of the subject device indicated that no abnormalities related to the reported phenomenon.From the above inspection result, it is concluded that the operation pipe was pushed and kinked while the coil sheath was sharply curved, and the operation pipe became bent and broken after the slider was pulled.The instruction manual of the subject device states below: always have the olympus loop cutter ((b)(4)) ready to cut the coil sheath, tube sheath and operation wire in case the loop cannot be detached from the instrument; do not withdraw the instrument or change the angulation of the endoscope before the look is detached from the instrument during a ligation.This report is being submitted as a mdr in an abundance of caution.
|