TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO TCM II COOLING AND HEATING SYSTEM Back to Search Results
Model Number 4415
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/29/2015
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the cooler heater unit was leaking water.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Per the field service rep ( fsr) verified the water leak from the rear panel of th unit.The fsr found the cardioplegia (cpg) pump outlet coupling was cracked.The fsr replaced the cracked coupling, as well as the cpg pump housing and 1/2" connector.The fsr tested operation and verified no water leaks.The unit operated the mfr specifications and was returned to clinical use.The suspect parts were returned to the mfr for further evaluation.If additional info becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO TCM II COOLING AND HEATING SYSTEM
Type of Device
TCM II COOLING AND HEATING SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4535926
MDR Text Key5516254
Report Number1828100-2015-00100
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/11/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/29/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-