• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problems Failure to Charge (1085); Failure to Conduct (1114); Use of Device Problem (1670)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Event Description
It was reported the patient last recharged her ipg in (b)(6) 2014 and the ipg is unable to communicate with her charging system.A st.Jude representative confirmed a spare programmer was unable to communicate with her ipg and displayed error message.It is unknown when the patient lost stimulation.Surgical intervention is pending to address this issue.
 
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Follow up information identified the patient underwent surgical intervention on (b)(6) 2016 to remove and replace her ipg which resolved the issue.
 
Manufacturer Narrative
(b)(4).
 
Event Description
Further information identified the patient experienced pocket heating while recharging.All issues were resolved with explant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4540392
MDR Text Key5508437
Report Number1627487-2015-26085
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/28/2013
Device Model Number3788
Device Lot Number3546773
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/22/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 1192 (2), SCS ANCHOR
Patient Outcome(s) Other;
-
-