The jet ventilator stopped working.An error message on the jet ventilator lit up: "ventilator fault".The patient's spo2 dropped to 86%.The patient's endotracheal tube (ett) was removed from jet ventilator and nurse (rn) began manually ventilating (bagging) the patient; the patient's sp02 returned to 100%.New jet ventilator obtained and the ett was transferd to the new jet ventilator. =
manufacturer response for jet ventilator, high frequency jet ventilator (per site reporter)
=
the jet ventilator was due for preventive maintenance in two weeks, under service contract.The ventilator was shipped back to manufacturer for service.Report received from manufacturer: "the whisper jet passed the self-test and operational verification".Under-went inspection, cleaning and calibration.Final validation after service: passed service operational verification and passed quality control inspection.Jet ventilator shipped back to our facility a month after inspection.
|