It was reported that the physician followed the instructions for use and inserted the iab-06840-u through the sheath via right femoral to the correct position, however the intra-aortic balloon (iab) could not fully open.The intra-aortic balloon pump (arrow kaat ii plus, s/n (b)(4)) alarmed "high pressure" and check for "kinked catheter," "partially wrapped balloon" and "balloon too large." as a result, the iab and sheath were removed together as one unit successfully.According to their judgment, the physician felt that the patient did not need the use of an iab and therefore did not insert another one for this case.There were no pump strips generated for review.The last x-ray was performed after placement and the findings were the iab could not fully open.The x-ray findings are available for review.There was no report of patient death, complications or injury.No medical/surgical intervention was required.There was less than a 30 minute delay or interruption in therapy with no harm to the patient noted.The patient outcome is the patient is currently in the intensive care unit.Additional information received on 02/13/2015 stated the user chose to use ecmo (extracorporeal membrane oxygenation) instead of iabp therapy.The patient outcome is fine now.
|