• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CODMAN & SHURTLEFF, INC. CODMAN DISPOSABLE PERFORATOR; DRILLS, BURRS, TREPHINES & ACC.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN & SHURTLEFF, INC. CODMAN DISPOSABLE PERFORATOR; DRILLS, BURRS, TREPHINES & ACC. Back to Search Results
Catalog Number 26-1221
Device Problem Plunge (1462)
Patient Problem No Information (3190)
Event Date 02/11/2015
Event Type  Injury  
Event Description
I received a phone all this evening from neuro charge nurse at (b)(6).He wanted to let us know about an incident with a 14mm codman perforator which plunged into the dura apparently because its clutch didn't engage.The lot number for this device is lg002s, and he tells me they had a similar incident in recent months.On (b)(6) 2014 per rep, the event occurred during the surgery, did not delay the surgery for more than 30 minutes and the surgeon responded by changing to another perforator.The patient's condition after the event: "ok".
 
Manufacturer Narrative
(b)(4).Upon completion of the investigation, a follow up report will be filed.
 
Manufacturer Narrative
Upon completion of the investigation it was noted that the customer¿s complaint of "perforator which plunged into the dura apparently because its clutch didn't engage" was not verified.This perforator met the test method acceptance requirements; proper engagement and disengagement were achieved with every drill hole.There was no premature disengagement or erratic and poor cutting action.The device history records for the perforator was reviewed and all test and inspections associated with the assembly process met specification requirements.No root cause could be determined as no problem was identified.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
325 paramount drive
raynham MA 02767
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key4548263
MDR Text Key5527969
Report Number1226348-2015-10125
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number26-1221
Device Lot NumberLG002S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received03/26/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/06/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-